FLOMID-MX
Control unit for
electromagnetic
flowmeters
INSTRUCTIONS MANUAL
R-MI-Flomid-MX Rev.:2
English version
instrumentación
para fluidos
Vista de pagina 0
1 2 3 4 5 6 ... 19 20

Indice de contenidos

Pagina 1

FLOMID-MX Control unit for electromagnetic flowmeters INSTRUCTIONS MANUAL R-MI-Flomid-MX Rev.:2 English version in

Pagina 2

10 Once the minimum flow has been configured the set-up returns to the main menu showing flow rate set-up. Pressing the push button the next ite

Pagina 3 - 2 INSTALLATION INSTRUCTIONS

11 P u l s e w i d t h 6 5ms After the option is chosen, the set-up returns to the main menu. Pressing the push button the ne

Pagina 4

124.4.1 Output 1 mode In this screen we select output No 1 to act as an alarm, as an indication of pipe empty or as a batching control output.

Pagina 5

13If the alarm option has been selected then the screens for selecting the switching points will appear. O u t p u t 2 O N a t > 0 0

Pagina 6

144.6 Preferences P r ef er enc es In this submenu we can configure other parameters, such as the default working screen an

Pagina 7

15N o r m a l f l o w +I n v e r t e d f l o w 0 N o r m a l f l o w 0I n v e r t e d f l o w + The sign “+” means that the c

Pagina 8 - 4 SET-UP

16 To program the languages the use of the push buttons is slightly different from the normal use. 1. Push button - To move to the next cha

Pagina 9 - DN · 0,014137 = Q

17 By pressing the key , the state of the digital outputs will change. By pressing the key , the state of the frequency output will change. By pres

Pagina 10

187 STATUS INDICATION During the normal working mode, some symbols can be displayed on the screen which indicate the status of the instrument. I

Pagina 11

199 TECHNICAL CHARACTERISTICS 9.1 Power supply Supply voltage: Standard : 220 VAC 50/60Hz, 24 VDC On order : 240 VAC, 110 VAC, 24 V

Pagina 12

TABLE OF CONTENTS 1 INTRODUCTION ...

Pagina 13

20WARRANTY Tecfluid S.A. GUARANTEES ALL ITS PRODUCTS FOR A PERIOD OF 24 MONTHS, after consignment, against all defects in materials and workmanship.

Pagina 14

31 INTRODUCTION The Flomid-MX electronic control unit adapts to the different Flomid and Flomat series electromagnetic flow sensors. The microp

Pagina 15

4 - A mains switch must be provided to disconnect the equipment. This switch must be marked as the disconnecting device for the equipment and be w

Pagina 16

52.4 Remote partial reset connection Terminal 11 Anode. 12 Cathode. This input is for connecting a remote push button to reset the par

Pagina 17 - 6 TOTAL VOLUME COUNTER

6 2.6 Sensor connection When the sensor is separate from the electronics the sensor cable is connected at the rear of the instrument. The connectio

Pagina 18

73. Push button - To change the display between flow rate, total volume or partial volume. - Together with to enter in the set-up mode

Pagina 19 - 9 TECHNICAL CHARACTERISTICS

84 SET-UP Set-up mode can be accessed by pressing the and push buttons at the same time, to avoid unwanted changes due to accidental access.

Pagina 20

94.1.3 Measuring units In this screen the measuring units should be introduced as follows: M e a s ur i ng uni t s l / h The cur