3.6 Vibrations
Vibrations of the pipes should be avoided by anchoring the pipe before and after the
sensor.
The vibration level should be less than 2.2 G in the range of 20 -150 Hz according to IEC
068-2-34.
3.7 Magnetic fields
Strong magnetic fields close to the sensor should be avoided.
3.8 Temperature
In open air installations it is recommended to install a protection to avoid direct sun light on
the flowmeter.
With thermally insulated pipes DO NOT insulate the sensor. High temperatures can
damage it.